Default Jobo ATL-2300 chemical pumping error

Thanks Ralf

It will not be necessary as I have now fixed the fault having spoken to
a very helpful lady on the JOBO Analogue Support desk in Germany.

The fix is as follows

Connect machine to power, but leave switched off.

Press Enter, Start and Power on buttons simultaneously. This brings up
the Service Menu on which there are a great many things - 14 items I
think.

Scroll down to No 11- Fill sensor. Press Enter, then toggle using up or
down arrow to change from With to W/O. Press Enter.

Press Power Off button.

Repair Complete!!!!

I have just run a test with same conditions as last night and the
machine pumped all chemicals in all bottles through with no errors.
It does mean that I have to fill with correct volume of chemicals for
each run. However this was part of my routine anyway in order to
conserve E6 chemicals.

For those with ATL2X00 machines, this Service menu could be the place
to start when things go wrong. I shall look there first in preference
to experiencing similar hassle to yesterdays palaver. I only hope that
I haven't written this in haste!

Baxter

  #32  
Old March 28th 06, 04:46 PM posted to rec.photo.darkroom
external usenet poster
 
Posts: n/a
Default Jobo ATL-2300 chemical pumping error

Baxter wrote:

It will not be necessary as I have now fixed the fault having spoken to
a very helpful lady on the JOBO Analogue Support desk in Germany.


I know her. If only there were someone like her in every company.

Ralf

--
Ralf R. Radermacher - DL9KCG - Köln/Cologne, Germany
private homepage: http://www.fotoralf.de
manual cameras and photo galleries - updated Jan. 10, 2005
Contarex - Kiev 60 - Horizon 202 - P6 mount lenses
  #33  
Old March 28th 06, 05:03 PM posted to rec.photo.darkroom
external usenet poster
 
Posts: n/a
Default Jobo ATL-2300 chemical pumping error


"Baxter" wrote in message
ups.com...

Press Enter, Start and Power on buttons simultaneously. This brings up
the Service Menu on which there are a great many things - 14 items I
think.

Scroll down to No 11- Fill sensor. Press Enter, then toggle using up or
down arrow to change from With to W/O. Press Enter.


Great ... that is the way it is done on more modern equipment with
non-volatile
memory.

See if there is a diagnostics section in the service menu ... if Jobo did
the software correctly it should be possible to display the status of the
liquid fill sensor on the display, this lets you fiddle with it and find
what it is that is causing the signal to drop.


 




Thread Tools
Display Modes

Posting Rules
You may not post new threads
You may not post replies
You may not post attachments
You may not edit your posts

vB code is On
Smilies are On
[IMG] code is Off
HTML code is Off
Forum Jump

Similar Threads
Thread Thread Starter Forum Replies Last Post
Kodak LS443 Camera Error E45 [email protected] Digital Photography 1 January 30th 06 05:02 PM
Road ruts with Jobo Brian Kosoff In The Darkroom 64 January 27th 04 01:08 AM


All times are GMT +1. The time now is 03:16 PM.


Powered by vBulletin® Version 3.6.4
Copyright ©2000 - 2024, Jelsoft Enterprises Ltd.
Copyright ©2004-2024 PhotoBanter.com.
The comments are property of their posters.